What is MARD?

MARD (mean absolute relative difference) is a measure of the accuracy of your CGM system.
If you're using or considering a Continuous Glucose Monitoring (CGM) system, you may have come across the term Mean Absolute Relative Difference (MARD). But what does it mean for you? Simply put, MARD is a way to measure how accurate your CGM is — the lower the percentage, the closer your sensor’s readings are to your actual blood glucose levels.1

How is MARD calculated?

Calculating MARD means checking how accurate your glucose monitor is. The accuracy is measured by comparing a series of blood sugar readings from your monitoring device to a trusted reference method, typically a laboratory blood sugar test or finger-stick blood glucose test. The difference between the reference value and the CGM reading is calculated. Then, the difference between the CGM value and the reference is turned into a percentage. Taking the average of those percentages will give the Mean Absolute Relative Difference (MARD).

Image
(left) BGM spot chart is compared to the curve char of CGMs. (right) Chart depicting the differences between BGM and CGM measurements, showing the difference between the BGM and CGM value, depicting the mean.
Example:
  • Pair the Accu-Chek SmartGuide sensor and app

    If your CGM reads 8.3 mmol/L and your BGM or lab test shows 8.9 mmol/L, the difference is 0.6 mmol/L. Across multiple tests, this difference value is calculated as a percentage, and then averaged into a value that represents your CGM’s accuracy.

What affects MARD results?

While MARD is useful, several factors can impact MARD values2:

Blood sugar levels:

MARD can vary across different glucose ranges. Accuracy may be better in mid-range glucose values and slightly lower in very high or very low readings.

Lag Time:

Because CGMs measure glucose in interstitial fluid rather than blood, readings may slightly lag behind actual blood glucose changes, especially after meals or during exercise.

Sensor wear time:

Some sensors lose accuracy over time, with slight changes in MARD values toward the end of their wear period.

Calibration requirements:

CGMs may require manual calibration with a BGM, and rely on the accuracy of those reference readings, which can influence MARD.

Individual differences:

Certain medications and external factors (such as vitamin C, aspirin, and dehydration) may affect sensor readings, causing fluctuations in MARD results. Skin thickness, hydration levels, and sensor placement can all impact how accurately a CGM reads glucose.

What MARD means for CGM users

MARD helps you measure your confidence in your CGM readings. However, regardless of the MARD value,it’s important to:

  • Always interpret your CGM data in context — consider trends, symptoms, and your daily routine.
  • Be mindful of lag time and discuss timing when making treatment decisions with your healthcare provider.
  • Confirm your values with finger stick tests when needed, especially if a reading conflicts with symptoms.

MARD is a helpful measure of CGM accuracy, but it’s not the only factor to consider. Although lower MARD scores mean more accurate readings, understanding your CGM data is key to maximizing its benefits. Consult your healthcare provider to find the most suitable glucose monitoring system for your diabetes care.

Next, learn more about Lag time.

Continue training

1. Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring. Diabetes Technol Ther. 2015 Mar;17(3):177-86. doi: 10.1089/dia.2014.0272. Epub 2014 Dec 1. PMID: 25436913; PMCID: PMC4346608.

2. Longo RR, Joshi R. The Devil Is in the Details: Use, Limitations, and Implementation of Continuous Glucose Monitoring in the Inpatient Setting. Diabetes Spectrum. 2022 Nov 1;35(4):405–19.